SEA6000VX HSFinder Analysis Examples

Film Thickness Measurement Example (1): Sn-Ag/Cu and Au/Ni/Cu

The SEA6000VX HS Finder also measures coating thicknesses typical of the SFT series applications. Application possibilities include ultra-thin Au coating measurement, also Pb (lead) hazardous substance analysis and simultaneous measurement of that coating thickness. For example, composition measurement of hazardous substances in Pb-free solder coatings, Sn coatings on lead frames, and electroless Ni coatings.
[Analysis of Sn-Ag/Cu]
Area: 1.2mm square
Film FP method Time: 100 sec (Pb filter)
  30 sec (no filter)
  [Au/Ni/Cu coating thickness]
Area: 0.2mm square
Cal method
Time: 30 seconds
Thick
Comp ratio
Haz Sub
 
Sn-Ag [µm]
Ag [wt%]
Pb[ppm]
1
5.34
2.83
291
2
5.30
2.63
297
3
5.31
2.64
347
4
5.34
2.73
303
5
5.31
2.67
329
6
5.33
2.67
334
7
5.33
2.75
291
8
5.30
2.74
262
9
5.35
2.72
303
10
5.36
2.91
308
AVE.
5.33
2.73
306.5
SD
0.02
0.09
24.6
CV
0.4%
3.2%
8.0%
     
 
Au [µm]
Nl [ µm]
1
0.054
5.11
2
0.055
5.11
3
0.054
5.12
4
0.055
5.08
5
0.054
5.10
6
0.055
5.13
7
0.053
5.08
8
0.053
5.09
9
0.054
5.08
10
0.055
5.14
AVE.
0.054
5.14
SD
0.008
0.02
CV
1.5%
0.4%

To Top close